FIBA TENS – 24-CHANNEL NEUROTHERAPY STIMULATION SYSTEM

Advanced Multi-Channel TENS/NMES for Precision Neuro Rehabilitation

CATEGORY

  • Medical Devices
  • Neurotechnology
  • Rehabilitation Tech

Confidential Prototype

SERVICES

  • Feasibility & Product Design
  • Full-Cycle Product Development
  • Hardware & PCB Engineering
  • Mechanical Engineering
  • Firmware Development
  • Prototyping (EVT, DVT, PVT)

PROJECT TIMELINE

JAN 2023 – AUG 2023

THE CHALLENGE

Deliver a portable, 24-independent-channel TENS/NMES device capable of targeting multiple muscle or nerve groups simultaneously, with: High-precision current delivery (0–100 mA per channel) and programmable waveforms; Complete isolation between channels to prevent crosstalk and ensure patient safety; User-friendly interface for clinicians and patients to configure up to 24 zones; Closed-loop adaptability, optionally integrating EMG feedback for automated parameter tuning; Compact, wearable form factor with all-day battery life and robust biocompatible enclosure.

  • High-Precision Current Delivery: 0–100 mA per channel, programmable waveforms
  • Channel Isolation: Complete isolation between channels for safety
  • User-Friendly Interface: Configure up to 24 zones easily
  • Closed-Loop Adaptability: Optional EMG feedback for automated tuning
  • Wearable Form Factor: Compact, all-day battery, robust enclosure

KEY HURDLES

  • •Minimizing channel interference while packing 24 stimulators into a handheld unit
  • •Designing an intuitive UI for managing dozens of stimulation protocols
  • •Keeping power consumption low enough for 8+ hours runtime on a single charge
  • •Ensuring medical-grade isolation and safety compliance across all channels

THE OPPORTUNITY

Current TENS/NMES systems usually cap at 4–8 channels, requiring multiple devices or limited therapy. FiBa TENS unlocks: Simultaneous multi-site therapy for complex neuro-rehabilitation regimens; Home-based neurotherapy, reducing clinic visits and improving patient adherence; Data-driven insights via cloud-sync of session logs, enabling remote clinician oversight; Rapid protocol switching to address spasticity, pain, or motor re-education in one session.

  • Simultaneous multi-site therapy for complex neuro-rehabilitation regimens.
  • Home-based neurotherapy, reducing clinic visits and improving patient adherence.
  • Data-driven insights via cloud-sync of session logs, enabling remote clinician oversight.
  • Rapid protocol switching to address spasticity, pain, or motor re-education in one session.

FEATURE SETS

  • •24 Independent Channels – Each channel delivers up to 100 mA with ±1 µA resolution
  • •Custom Waveform Library – Monophasic, biphasic, burst, Russian current & more
  • •EMG Feedback Option – Auto-adjust intensity based on real-time muscle response
  • •7″ Capacitive Touchscreen – Configure channels, view live current/voltage graphs
  • •Wi-Fi & BLE – Sync therapy logs to cloud portal or mobile app
  • •12-Hour Battery Life – 10 000 mAh Li-ion pack with smart power management
  • •Biocompatible Enclosure – ABS+PC overmold, IP54 ingress protection
  • •Safety Suite – IEC 60601-1 compliance, per-channel over-current & isolation monitoring
  • •32 GB Local Storage – Save thousands of session profiles and waveforms
FIBA TENS Product Photo

GALLERY

Front Panel – Touchscreen UI showing channel matrix

THE SOLUTION

Axio Electronics delivered a portable, multi-channel neurotherapy system with advanced isolation, real-time analytics, and clinician-friendly design.

24 Independent Stimulation

24 Independent Stimulation

Fully isolated current sources, up to 100 mA per channel

WAVES

WAVES

Programmable Waveforms – Library of clinical protocols plus custom waveform designer

EMG

EMG

Feedback-Driven Control – Optional EMG integration auto-adjusts intensity in real time

UI

UI

7″ Touchscreen Interface – Channel grouping, real-time graphs, and quick-start presets

NET

NET

Wi-Fi & BLE Connectivity – Secure TLS sync to cloud portal; BLE link for mobile app

BATT

BATT

12-Hour Runtime – High-capacity pack with dynamic load balancing

SAFE

SAFE

Full Medical-Grade Safety – Per-channel fault detection, isolation barriers, IEC 60601-1

LOG

LOG

32 GB Onboard Storage – Archive thousands of session logs and waveform captures

OUR APPROACH

Mechanical Engineering Process

Mechanical Engineering Strategy

  • Stacked Modular Layout: Separate housings for power, stimulation, and UI to simplify isolation

  • Rapid-Swap Battery Module: Tool-free replacement for uninterrupted multi-session use

  • Wearable Mount Options: Belt clip and tabletop cradle for clinician or patient use

  • Ingress Protection: Gasketed seams and sealed connectors for IP54 rating

Mechanical Engineering Process

Mechanical Engineering Process

  • Finite-element analysis for heat dissipation in high-density PCB stack

  • Rapid 3D-printed mockups to validate electrode-lead ergonomics

  • Drop and ingress trials ensuring robustness in clinical environments

  • Final tooling for overmold and injection-mold parts

PCB Design

PCB & Hardware Design

  • 24 Current Driver ASICs with galvanic isolation per channel

  • Dual-Core MCU (ARM Cortex-M4 + DSP) for waveform generation and safety monitoring

  • EMG Front-End: Low-noise amplifiers and 16-bit ADC for optional feedback mode

  • Power Management: High-efficiency buck converters and isolation transformers

  • Radio Modules: Wi-Fi 802.11ac and BLE 5.2 with separate antennas for reliability

Firmware Design

  • Real-Time Waveform Scheduler: Sub-millisecond switching across 24 channels

  • Safety FSM: Hardware watchdogs and per-channel over-current shutdown

  • EMG Control Loop: Adaptive PID adjusting output to maintain target muscle activation

  • UI Framework: Qt-Embedded interface optimized for gloved-hand operation

  • Connectivity Stack: MQTT over TLS for secure cloud logging; GATT profile for BLE control

  • OTA Updates: Verified firmware bundles delivered via Wi-Fi or BLE

Firmware Development

Prototyping & Validation

  • EVT: Bench-tested all 24 channels for isolation, linearity, and waveform accuracy

  • DVT: Clinical pilot with 10 patients—validated multi-site therapy efficacy and usability

  • PVT: 50-unit pre-production run—automated jig tests for channel calibration, UI responsiveness, and battery endurance

In-House Prototyping Process

Key Challenges & Resolutions

  • Channel Crosstalk
    — Added high-speed analog isolators and optimized PCB ground planes
  • Thermal Buildup
    — Integrated copper heat spreaders and active venting in housing
  • UI Complexity
    — Introduced channel-group presets and quick-start templates
  • Battery Swell
    — Upgraded to pouch cells with pressure relief vents
Challenges Illustration

PILOT & PRODUCTION

PILOT & PRODUCTION

After completing engineering validation and multi-phase bench testing, FIBA TENS entered clinical pilot trials and pre-production refinement to validate its 24-channel neurostimulation capabilities in real-world therapy sessions:

  • Clinical Pilot: Deployed in two neuro-rehabilitation centers with 10 patients; therapists reported improved session efficiency and simultaneous multi-muscle stimulation across 12+ sites.

  • Design Refinement: Reworked electrode lead harness and overmolded pod grip after pilot feedback to enhance usability, durability, and patient comfort during extended sessions.

  • BOM Optimization: Reduced BOM by 11% through consolidation of stimulation channels onto denser ASICs and switching to higher-efficiency power modules.

  • Certifications: Designed for IEC 60601-1 compliance; internal safety testing passed per-channel overcurrent protection, isolation, and leakage standards. Final certification pending production batch.

  • Manufacturing Test Fixture: Developed automated jig to verify stimulation waveform accuracy, EMG input response, UI latency, and battery runtime across all 24 channels before shipment.

Rocket Illustration

How It Works

1.

Connect Electrode Pods to the 24-lead harness and attach to target sites.

2.

Power On and log in via touchscreen or mobile app.

3.

Select Protocol: Choose pre-configured therapy or design custom waveforms.

4.

Start Session: Live current and voltage graphs update; EMG feedback (if enabled) auto-tunes.

5.

Sync Data: Completed sessions upload automatically for remote clinician review.

WHAT CAN
WE CREATE
TOGETHER

DOWNLOAD THE FIBA TENS CASE STUDY (PDF)

Get a PDF version of the FIBA TENS case study to download and share.

Client Testimonial

np8i
user-image.png

Philip Muiico

AxioBionics LLC

The Axio team supported us for several years as our R&D partner across multiple prototypes, including a few patented technologies. Their primary contributions were EMG and TENS stimulators. We were consistently impressed by their exceptional talent, advanced R&D tools, and systematic approach.

Axio ElectronicsAxio Electronics

Axio Electronics provides end-to-end electronic design and manufacturing services across industries, from consumer electronics to high-tech startups. We prioritize understanding customer needs and delivering top-tier engineering for design, validation, testing, and production.

CONTACT

LOCATION: 520, JTM, MODEL TOWN, JAIPUR, RAJASTHAN, INDIA - 302017

EMAIL: ramprakash@axioelectronics.com

PHONE: +91 9079253395

AXIO ELECTRONICS © 2025 | PRIVACY POLICY
WhatsApp